Temp. (correction of temperature)
If a temperature correction has been set for the calibration (
input field labelled Temp. are shown above the graphic.
1. You can now manually change the temperature correction values here in one of the following
ways:
– Moving the slide control
– Manually entering a value
– Pressing the arrow fields
The curves delimiting the correction range are moved on the chart.
Graphic > Settings
You can display a larger or smaller view of the chart. To call up the dialog for the settings
described below, click on one of the two axes in the graphical depiction.
auto
1. Select this box to enable the automatic adjustment of the view.
Activated: The values for the X and Y axes are automatically adjusted. The zoom
factor is set to 100%.
Deactivated: Enter the values yourself.
Zoom
2. Enter any value between 1 and 400.
X axis/Y axis
3. Enter the lowest (Min) and highest (Max) value for the display area.
Activate temperature/density correction reference lines
Activate
O
Note: You can activate temperature or density correction.
Activated: Measurement results are corrected accordingly if fluctuating measurement
temperatures are detected.
Deactivated: No temperature correction takes place.
Increment
1. Enter a temperature correction factor.
2. Click Apply to accept the settings.
Graphic > Print
The graphical depiction of the calibration can be printed.
See page 53 for information on how to connect a printer.
Select Print > from the Calibration > Moisture calibration > Graphic
menu.
The graph is printed as it appears on the screen.
90