In so-called multiplex units with 2 sensors attached, the system loads the modes supported by
both sensors, i.e., the number of available modes increases.
O
Note: Imported sensor modes can neither be edited nor deleted.
In case the sensor modes were not automatically loaded when the system booted, you can also
choose System settings > Configuration > Sensor modes > Read from the menu and load the
modes later.
Editing Sensor modes
Prerequisites:
You are logged on and own level 4 or 5 access rights.
There are basically two cases in which it is necessary to edit the sensor modes:
You are using a sensor of an older type or, for any other reason, your sensor is not connected to
port Sensor 1/2 of your LMA3xxPA / PMD3xxPA. In this case, the sensor modes will not be rec-
ognized automatically and you will have to manually define them yourself using the sensors' data
sheet as a basis.
Choose System settings > Configuration > Sensor modes > Edit > Create and transfer the
information from the data sheet to the Sensor mode dialog fields. Press Apply to have the new
sensor mode available for measuring products.
Requirements have changed because the sensor has been modified (e.g. due to dirt) or because
of new or modified samples not covered by the existing modes. In this case, you would copy an
existing sensor mode and adapt the copy as appropriate.
1. Choose System settings > Configuration > Sensor modes > Edit > Copy.
2. Pick a sensor mode from the table that is suitable to act as a basis for the new sensor mode; then
press Copy. A copy of the sensor mode will be added to the list under the same name as the
original mode.
3. Choose System settings > Configuration > Sensor modes > Edit > Values.
4. Go down the list, select the new (copied) sensor mode and press Values.
5. On the Sensor mode page, adapt the parameters as appropriate. In order to avoid confusion,
you are strictly recommended to change the Name of the sensor mode. Press Apply to have the
new sensor mode available for measuring products.
Which of the details need to be changed and in what way, depends on the how the requirements
have changed.
O
Example: When measuring a product of higher moisture or density, extend the search range by
a lower Start frequency value.
6. Press on Library to be shown a list of the common sensors in a new dialog.
7. Pick the name of the connected sensor from the list.
In some cases, diverse modes are shown for one sensor (lo, med, hi).
LMA316PA / LMA321PA - Other functions