Note: Press the navigation controls (see Navigation (Details on page 72)) to view products you
cannot see immediately.
3. Press Apply. The calibration for the selected product can now be created or edited.
Moisture calibration
See Calibration (Details on page 35) for a step-by-step description of the calibration procedure.
Calibration in the laboratory
The table is called up for measuring the reference values and for entering the laboratory values.
Table
Select Calibration > Moisture calibration > Table to go to the calibration table view. This view
is for measuring reference values, entering laboratory values, and editing the calibration.
1. Press the Calibration - Moisture mode display or the table symbol to switch to the table view.
This menu has the following sub-items:
– Table > Edit (Details on page 85)
– Table > Settings (Details on page 88)
– Table > Print (Details on page 88)
– Table > Export (Details on page 89)
O
Note: If the table contains more than 10 values, use the navigation controls. See Navigation
(Details on page 72).
Table > Edit
Values
The reference values for the measured samples are entered into the calibration table here.
1. In the menu, select Calibration > Moisture calibration > Table >
Edit > Values.
In the submenu, select whether an individual entry (Remark, Reference moisture, Type etc.) or
all values are to be edited:
– Remark (Details on page 86)
– Reference moisture (Details on page 86)
– Type (Details on page 86)
– All values (Details on page 87)
The table is now displayed in edit mode, i.e. the controls for selecting one or more rows
(
page 70) are displayed at the bottom.
1. Press a row to select it.
LMA316PA / LMA321PA - Operator Interface - Reference
85