118
LMA316PA / LMA321PA - Operator Interface - Reference
Explanation
The empty sensor adjust is carried out automatically for
process measuring devices if the sensor is empty. The
interval to be used for carrying out empty sensor adjusts
can be defined in Measurement interval.
Specifications on the sending of measurement data via the
serial interface or Ethernet TCP/IP.
In the case of the PMD320PA, you can define here whether
you want the density to be displayed on the screen, too.
After each save cycle, the system checks the age of the last
empty sensor adjust measurement and issues a warning
(i.e. warning indicators light up) if the specified value (in
days) has been exceeded.
Defines the required precision (stability) of the empty meas-
urement. If this stability is not achieved, the measurement
is terminated with an empty sensor check error. In gen-
eral 0.01 MHz is an optimum value. This applies to all
empty measurements including those that are carried out
automatically during a laboratory or process measurement
sequence.
Optionally, you can hide the average values from the meas-
ured values with a 24 V digital input.
Defines the type of process measuring device.
Moving average values for moisture, density, temperature
etc. are calculated and periodically stored for all running
measurements using the time defined here. The interval for
saving the data is defined in Measurement interval.
Time period for which all measured values are determined
and copied to the calibration after the Sampling button has
been pressed.
Available time period before the empty measurement for
removing the product/inserting an empty measuring beaker
after the appearance of the Please empty! message.
Available time period before the measurement for filling the
sensor after the appearance of the Please fill! message.
Available time period for saving the measurement results
after the measurement.
Time span for inserting temperature sensor Pt 100 into the
sample. If using IR sensors, no pause is required, i.e. the
value can be set to No.
Maximum time that the system waits to see if the stability
criteria defined in Global settings/Temperature are met for
the temperature measurement (
page 109).