13.5
Calibration
A calibration must be carried out once for every product.
Every calibration begins with the selection of a product. See Product choice (Details
on page 84).
Choose whether to carry out a moisture or density calibration. See Moisture calibration (Details
on page 85) or Density calibration (Details on page 94).
Please note that the procedure for calibration in the laboratory is different to the procedure for the
in-process calibration.
A single calibration is assigned to every product. This calibration can be enhanced and refined
over time. The calibration table with all entered values is saved automatically.
Product choice
You call up the product choice dialog using the menu or by pressing the product names.
To create new products, choose Product settings (Details on page 103).
O
Note: The currently selected product is indicated by a green tick in the No. column.
O
Note: The REFERENCE product is only required for the optional "reference standard" function.
1. Press the column header to sort the table. The current sort order is indicated by an arrow in the
column header.
2. Press the row containing the name of the product to be calibrated.
O
Note: Products with similar properties can reference the same calibration. If a product references
the calibration of a similar product, corresponding information can be found in the Reference to
calibration line. In the case of products with their own calibration, the Cal. link column contains
the entry self.