Sartorius LMA316PA Betriebsanleitung Seite 68

Mikrowellen-feuchtemessgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

WLM6009-a10081
1. If using a device with a touchscreen, press the screen to go to the Measurement view at any
time.
O
Note: Users who are only carrying out measurement tasks simply follow the instructions shown
here.
O
Note: Some devices which are used for process measurement only have a display screen which
either shows the measured values or, if appropriate, an error code.
Display panel 1 / 2 (Measurement)
Two display panels are available here. Each of the two displays can be set individually.
The display screen shows one of the following items:
– The currently measured value/value at the cursor position for moisture, density, or temperature
OR
– The statistics for selected measurement values or all measured values
1. Press one of the display panels to change the settings for the display. See section Analogue dis-
play (Details on page 81).
Select Measurements > 1st/2nd display from the menu to change the type of the display.
Display panel 1 / 2 (Calibration)
During calibration, the display panels show information on the calibration.
Type:
The calibration type is displayed here (linear or non-linear and subdivided into segments where
appropriate).
Values:
This shows the following:
The number of calibration points
Whether the calibration is computed automatically and whether the segment limits are automati-
cally recalculated where applicable
At least two calibration readings and the corresponding reference values are required to calculate
a calibration.
1. Press the Calibration display panel to change the calibration settings. See section Type set-
tings (Details on page 91).
68
LMA316PA / LMA321PA - Operator Interface - Reference