Note: Temperature and density variations can only be taken into account if corresponding
values are available for the samples.
As a rule, the temperature should be taken into account in the case of temperature variations of
more than ±5°C.
3. Press Apply to accept the setting and calculate the calibration.
Once the calibration has been successfully calculated, the parameters are displayed in the
1st/2nd display panel. See 1st display/2nd display.
4. Press Expert to make all settings manually .
5. Press auto to calculate the calibration automatically.
Expert
You can specify the calibration settings manually here. Settings made beforehand (linear/non-
linear, number of segments) are applied but can be changed here.
You can set the limits of the individual segments of the calibration curve.
You can modify the correction value manually. If you activate this option for temperature or
density, you can use the slider control or number input function in the graphical depiction of the
calibration curve to set the correction value.
Results
The right-hand side of the dialog shows the results of the settings made. These fields cannot be
edited.
If the calibration cannot be calculated using these parameters, a message to this effect is
displayed in red at the bottom of the Results panel.
Press Calibrations to display a list of all possible calibrations. This list is displayed in the same
dialog on the second Calibrations tab.
The list of calibrations can be filtered to improve clarity. Press the All calibrations list field and
select a filter.
Press an entry to select it.
Press Choice to apply these parameters for the calibration.
Press Display to return to the graphical depiction or table view.
In the graphical depiction or table view, press Save to save the calibration or return to the dialog
with the expert settings.
O
Note: If using the program on your PC, press Apply to adopt the setting.
Press Simple to return to a simple calibration process.
Press List to display the parameters for the chosen calibration in a list.
auto
The calibration is calculated automatically. If you want to change the settings for calculating the
calibration, press the 1st display panel (calibration, type, correction). The dialog for the simple
calculation of the calibration (
92
LMA316PA / LMA321PA - Operator Interface - Reference
page 91) appears.