10. Service and maintenance

10.1
Software update
O
Note: If you are controlling the measuring device using an external PC, you must also update the
software on the external PC following a software update on the measuring device.
You receive software updates from Sartorius as zip files.
1. Create a new, empty folder on your PC and unzip all files from the zip file. Make sure that you
retain file paths and the folder structure of the files.
2. Copy the unzipped files and folders to the USB stick you received with your measuring instru-
ment.
3. Turn off the device using the on/off switch at the back.
4. Insert the USB stick into one of the USB ports.
5. Turn on the unit using the on/off switch at the back.
The update is automatically detected and installed.
O
Note: If the device does not seem to react during the startup process, remove the USB stick and
then insert it into a USB port again.
10.2
Creating a backup/restoring the system
Creating a backup
When you create a backup, all of the device settings including the calibrations and measured
values are stored on the connected USB stick or PC:
You should do this regularly since it means that you can restore a recent state if problems occur.
It also allows the archiving of old measured values.
The Backup function can only be used by users with level 4 access rights or higher. If the meas-
uring device is currently being used with a login assigned to another level, you must log in again
with an appropriate user as follows:
1. Select the Change user menu item.
2. When the Login dialog appears, enter your user name and password.
3. Click Apply. This releases the functions in the Service menu.
O
Note: After carrying out administrative work, you should either change your login to your "normal"
access rights or, if a default user is defined, restart the control software to prevent unauthorized
modifications of the system settings or calibrations.
Your next steps differ depending on whether you are working directly on the measuring device
with a USB stick or are controlling the device using an external PC.
LMA316PA / LMA321PA
59