WLM6009-a10081
Statistics
Statistics can be created for selected entries in the table.
Once you press Statistics, corresponding buttons become available. See Controls in
Measurement/Calibration (Details on page 70).
1. Select an entry, a block, or all entries.
The same remark is defined for all selected entries.
2. To select several rows, press Multi.
Now select several values using the Block 1/Block 2 control or select entries individually.
3. Press Statistics.
Assuming that the corresponding settings have been made, the statistics are depicted in one of
the display panels at the top right of the screen. See Statistics (Details on page 83) to learn more
about how to configure the display panels.
4. Press Close to return to the display.
Graph
The measured values are depicted as a graph.
The graph visually illustrates extreme values and shows trends in the development of values.
You can choose whether to display the position or time of measurements on the horizontal axis
(X axis).
You can display one or two curves, with a choice of moisture, density, and temperature.
You can make these settings and more by choosing Measurement > Graph > Settings (Details
on page 78).
1. To start or stop the measurement process and to take samples and carry out an empty sensor
adjustment, use the controls described in Measurement/Calibration (Details on page 70).
Determining the cursor position
The values of the measurement at the cursor position can be displayed in one of the display
panels (Display panel 1 / 2 (Measurement) (Details on page 68)).
2. Press the graph. The cursor is fixed to this position. When you switch to the table display
(
page 74), the cursor position is retained.
3. Use the controls described in Navigation (Details on page 72) to set the cursor to a particular
position or to display further values in the measurement series.
O
Note: Press the Goto last control once.
The control remains activated, and therefore has a blue background. The cursor now remains at
the position of the last current measurement. The graph is constantly refreshed.
4. Press the Measurement mode display or the graph symbol to switch to the readings display.
Use the Zoom in and Zoom out controls to reveal more (zoom in) or less (zoom out) details in
the graph.
The zoomed in view displays the curve with the cursor at the centre of the visible X axis. If you
move the cursor to a new position by pressing the graph, the display of the current section is
moved accordingly.
A bar display above the graph clarifies which section of the overall curve is currently being
displayed. If an empty rectangle is shown, the entire curve for the whole measurement series is
displayed.