13.3
Controls
Measurement/Calibration
These controls are used during measurements and calibrations.
Only controls which are relevant to the task being carried out can be used.
70
LMA316PA / LMA321PA - Operator Interface - Reference
Start
Start the measurement process.
This starts the reference measurement during calibration.
Stop
Stop the measurement process.
Store
Saves the currently displayed value
ES adjust (empty sensor adjust)
Press this control to confirm that the sensor is presently empty.
You will be asked whether you want to perform a hard or soft empty sensor
adjustment.
Soft adjustment:
You simply inform the device that the sensor is empty or that there is an empty
measuring cup in the sensor.
Hard adjustment:
A hard adjustment is recommended if measuring results seem inaccurate or
distorted after relatively long periods of measuring. Again, you inform the de-
vice that the sensor is empty. At the same time, however, some internal pa-
rameters are adjusted so that material wear on the sensor or measuring cup
is taken into account. A new absolute empty reference point is set.
To perform an empty sensor adjustment, you need to be logged on with appro-
priate rights.
Reference standard
This function is designed to carry out a technical check of the device on the basis
of a reference body which can be acquired as an optional extra.
For more information, see page 57.
To execute this function, you must be logged onto the software with the appro-
priate rights.