Control via network PC
1. Connect the PC and the device to an existing IP network using network cables.
2. If not already noted, find out the IP address of the measuring device , see Determining the IP
address of the measuring device (Details on page 21).
O
Note: When incorporating the device into an already existing network, it is not always possible to
use the factory-set standard IP address, see also Changing the IP address of the measuring
device (Details on page 21).
3. Install TMV on the PC (see page 17).
4. Run TMV on the PC.
5. Click on the button with the house icon and choose System Settings > Configuration > Com-
munication > MW Devices.
6. Create a new device entry by choosing System Settings > Configuration > Communication >
MW Devices > Edit > New:
Choose the type MW4xxx.
Choose the interface TCP/IP.
Enter the IP address of the measuring device.
7. Click on Connect.
Controlling several measuring devices by one PC
If you intend to control several measuring devices via one PC, you need to create a respective
entry for each device. These measuring devices are to be allocated different IP addresses, see
Changing the IP address of the measuring device (Details on page 21).
You can then run the TMV software several times, connecting it to a different measuring device
each time, thereby enabling parallel processing on all measuring devices.
Determining the IP address of the measuring device
To determine the IP address of the measuring device proceed as follows:
1. Select System settings > Configuration > Communication > This TMV on the device. The
This TMV dialog for this measuring device is displayed and you can read the IP address in the
Address field.
Changing the IP address of the measuring device
If you intend to add the measuring device to an existing network, the measuring device is to be
allocated a suitable network address. Generally, the measuring device's IP address is set by the
service personnel via the service interface at installation in accordance with your demands.
If, however, the measuring device was reset to factory settings, (
located its standard address 172.28.1.1. Network access is no longer possible. The measuring
device must be re-assigned the desired address via direct connection.
For measuring devices with touchscreen (LMA321PA) or external monitor and keyboard
(LMA316PA), this change can be implemented directly on the device:
1. Select System settings > Configuration > Communication > This TMV on the device.
2. Enter the IP address in the This TMV dialog and state further network connection details, if appli-
cable. Contact your network administrator for further information on your network.
3. Finally, press on Apply.