– Define the measurement mode:
For laboratory measurements, deactivate the continuously option to enable the carrying out of
individual measurements as described in steps 7 to 10.
6. Perform an Empty sensor adjust.
7. Slowly fill the sensor until the header line shows a measuring value.
8. Press Start to store the value.
9. Slowly continue filling the sensor and press Start again to store the measured value.
10. Repeat this procedure until the sensor is completely full.
The results measured in the test are displayed in a table.
You can depict individual points on the plotted curve using different symbols and you can delete
points from the curve if required.
11. Press the line showing the operating mode (Compression test) to switch to the graphical depic-
tion.
O
Note: During the compression test, the 1st display/2nd display panels show important test
parameters:
Measurement settings
You use the Measurement settings dialog to change the parameters for the measurement
procedure.
1. Select Service > MW test > Compression test > Measurement settings or press 2nd display.
2. Activate the continuously option if you want to take compression test readings every second
(process measuring devices). Deactivate this option if you want to trigger and save each meas-
urement separately (laboratory devices).
3. Activate the overwriting option to overwrite old measured values once the memory capacity
reserved for the test is exceeded.
Memory
Specify the amount of memory to be reserved for the test here.
Current type
Determine the way the measured values are depicted in the graph here.
Settings and evaluation of results
To evaluate test results, select the graphical depiction of values.
1. Press the graph to display the Compression test properties (Details on page 34) dialog.
LMA316PA / LMA321PA - Calibration
1st display panel:
2nd display panel:
Compression test
Measurement set-
properties
tings