O
Note: Values are displayed rounded off in accordance with the settings in Global settings
(Details on page 108). When making entries, enter all decimal places.
3. Press Close. The input process is complete.
Remark
The Text input (Details on page 124) dialog appears. Remarks are free-text entries.
1. Press Apply to add the entry to the table.
O
Note: This dialog is displayed automatically following every measurement of a calibration value.
Typically, you enter the number of the sample here. This allows samples and measured values
to be clearly assigned to each other.
Reference density
The Numerical input (Details on page 125) dialog appears. Enter the laboratory value for the
density of the sample with all decimal places. The display of the values is rounded off in accord-
ance with the settings.
Press Apply to add the entry to the table.
This dialog is displayed automatically following every measurement of a calibration value.
Type
You can depict individual points on the calibration curve using different symbols. This allows you
to display measurements made on different days more clearly on the calibration curve, for
example.
1. Press an input field to display the Type dialog.
2. Select whether the calibration point is to be used for the calibration.
Valid
The calibration point is valid and is used to calculate the calibration curve.
The value is not taken into account for the calibration. However, the entry remains in the
table and can be set to Valid at a later time if required.
Symbol
3. Press the list field and select a symbol.
In the calibration graph, the calibration point is represented by the selected symbol.
4. Press Apply to accept the setting.
All values
The Calibration properties dialog appears.
Existing information is shown. Press an input field. The Numerical input (Details on page 125)
or Text input (Details on page 124) dialog appears.
If you selected the option for entering the type, the Type (
Press Apply to accept the setting.
LMA316PA / LMA321PA - Operator Interface - Reference