When delivered, the LMA3xxPA / PMD3xxPA normally only contains test items or pre-calibrated
items as well as one item - which cannot be edited - for the measurement of the reference body
(available as an option, see Reference standard (Details on page 57)).
Prerequisite
You are logged in with at least level 4 user rights (
1. Select the Product settings > Edit menu item.
2. Select the Create menu item.
3. Enter the following in the Product settings (Details on page 103) dialog:
General
No.
Enter a unique item number for each product.
Enter a unique product name to enable the clear identification of the product in the Name field.
If required, add an explanatory text in the Remark .
Cal. link
If a non-referenced calibration is being carried out for the product, choose self here.
If the calibration of a similar product is referenced, select the ID number of the product in
question.
Sensor mode
Select the sensor and the corresponding mode with which the product is measured. If the cor-
rect mode is not shown, you must manually define a mode, import one, or edit an existing
mode definition (
nected sensors.
O
Note: As a general rule, you should first choose a highly sensitive sensor mode. If this mode does
not return any results, reduce the sensitivity by selecting another, less sensitive mode.
Moisture/Density
Offset
Specify an offset if a similar product is referenced in the calibration but a certain standard de-
viation applies or if there is a constant difference between the value displayed and the actual
value (laboratory).
Upper limit /Lower limit
You can enter limit values here. If these values are exceeded by a measurement, an alarm
event is triggered.
Memory
Specify the amount of memory space to be reserved for results obtained from the product
here. The number you enter is equivalent to the number of rows in the table of measured val-
ues.
The memory value range is dependent on the number of products and calibrations - usually
around 2000 memory cells are left for measurements.
Select auto if a product is allowed to make complete use of the entire memory available for
the measured values.
If you are measuring multiple products, distribute the memory cells (around 2000) amongst
the products by entering appropriate values.
page 110). The device normally detects the modes provided by the con-
LMA316PA / LMA321PA - Calibration
page 30).