LMA316PA / LMA321PA - Operator Interface - Reference
Cause and corrective action
0
Measured values out of permitted range.
Perform a hard empty sensor adjust (
In the compression test properties, check the parameters in Measurement
range (MR)[Mhz]:
x = 9...Parameter A, min too high
x = 10...Parameter A, max too low
x = 11...Parameter B, min too high
x = 12...Parameter B, max too low
Call up the Compression test properties dialog by choosing Service > MW
test > Compression test > Graphics > Settings.
0
The floating average values are not in the permitted range (see 4x0).
0
The calibration cannot be calculated as there is no product temperature
(choose Calibration > Calibration - Moisture or Calibration - Density >
Type settings, Calibration type) for the calibration correction process.
Make sure that the product temperature sensor is connected and working
properly.
0
If you activated the Temperature coefficients option in the Sensor modes
dialog, you must connect a sensor that returns temperature values.
0
The moisture (or density) values exceed the given range.
Change the values in the product settings (see Product settings (Details
on page 103)).
page 70).