2. Press the control to start editing.
If you selected the option for entering a single value (reference moisture, reference density
etc.), the corresponding Numerical input (Details on page 125) or Text input (Details
on page 124) dialog is displayed.
If you selected the option for entering the type, the Type (
If you selected the option for entering all values, the Calibration properties (
is displayed.
O
Note: You can also press the relevant field in a table entry to enter a reference value, type, or
remark. If you do so, the input dialog is displayed.
O
Note: Values are displayed rounded off in accordance with the settings in Global settings
(Details on page 108). When making entries, enter all decimal places.
3. Press Close. The input process is complete.
Remark
The Text input (Details on page 124) dialog appears. Remarks are free-text entries.
1. Press Apply to add the entry to the table.
O
Note: This dialog is displayed following every measurement of a calibration value. Typically, you
enter the number of the sample here. This allows samples and measured values to be clearly
assigned to each other.
Reference moisture
The Numerical input (Details on page 125) dialog appears. Enter the laboratory value for the
moisture of the sample with all decimal places. The display of the values is rounded off in accord-
ance with the settings.
Press Apply to add the entry to the table.
This dialog is displayed following every measurement of a calibration value.
Type
You can depict individual points on the calibration curve using different symbols. This allows you
to display measurements made on different days more clearly on the calibration curve, for
example.
1. Press an input field to display the Type dialog.
2. Select whether the calibration point is to be used for the calibration.
Valid
The calibration point is valid and is used to calculate the calibration curve.
The value is not taken into account for the calibration. However, the entry remains in
the table and can be set to Valid at a later time if required.
86
LMA316PA / LMA321PA - Operator Interface - Reference
page 86) dialog is displayed.