9.4
Reference standard
This function is designed to check the technical functionality of the sensor and measuring device.
You may purchase a reference solid for your sensor, if you wish. This reference solid matches
the respective sensor and must only be used for the sensor for which it was delivered. The refer-
ence solid comes with a certificate listing the values of the solid.
Reference standard using TMV
1. Stop the current process to make sure that the sensor is empty, if appropriate. The sensor must
be free from product residues.
2. Select the product REFERENCE.
3. Perform an Empty sensor adjust.
4. Fill the reference solid into the sensor or place the reference solid on the sensor (for planar sen-
sors).
5. Perform the measurement(s) and compare these with the values stated on the certificate.
6. If the values exceed the limits stated on the certificate, contact the
Sartorius service.
LMA316PA / LMA321PA - Other functions
57