Type settings
(Type settings)
Here, you define the type of calibration.
Choose from the following:
Simple
Expert
auto
O
Note: You can also display this dialog by pressing the 1st display field (calibration, type, correc-
tions).
Simple
This dialog enables the simple calculation of the calibration.
1. From the list field, choose either the simple linear calibration type or the non-linear calibration
type, which - depending on the available values - can consist of up to four segments.
Linear: The calibration curve is a straight line.
Non-linear: The calibration curve is not necessarily a straight line. Areas of the calibration curve
may appear as curves.
O
Note: Please note that for a more complex calibration type more measured values are required,
as would be expected.
Extremum
In the case of the non-linear calibration type, you specify whether the curve is to have an extrem-
um here. Select the Extremum option if you know that one of the values represents a top or bot-
tom value.
Curve with extremum
Curve without extremum
Correction
2. Specify whether and how a correction is to be carried out for the calibration curve here. The cal-
ibration process then takes variations concerning the temperature and/or density of the samples
into account in the manner specified.
The system issues a corresponding message if there are insufficient measured values to enable
a sensible correction.