Moisture value min range
Specify the minimum difference between two values for those values to be valid for the calibra-
tion. Separate minimum differences are set for the measured calibration value, reference value,
correction value, and temperature correction value.
You can also call up this dialog by pressing the 2nd display panel (correlation, standard devia-
tion, values).
Automatically calculate
Activated: The curve is automatically recalculated each time the calibration values
change.
Deactivated: You have to click the 1st display panel again to recalculate the curve.
Deactivate this option if you still need to carry out considerable value editing. You are then
able to work more quickly since the calculation process always requires a certain amount
of time. If individual values are removed or changed, it may mean that the curve can no
longer be calculated. If this is the case, the system issues an error message.
With limits
Activated:
The limits of the non-linear zones are recalculated for every change.
Deactivated:
The limits of the non-linear zones are maintained when changes occur.
Optimized manual correction
Activated:
Measured temperatures are taken into account.
Deactivated:
Measured temperatures are not taken into account.
LMA316PA / LMA321PA - Operator Interface - Reference