Anhang | ASCII-Kommunikation
optoCONTROL 2520
Fehlermeldung
E37 ROI start must be lower than end.
E38 Too much output values for
RS422 enabled.
E41 The given passwords are not
equal.
E43 Software triggering is not active.
E46 Unsupported character.
E51 The given signal is not selected
for output.
E55 Synchronization and triggering at
level or edge are not possible at the
same time.
E56 Video signal is to dark.
E57 Measuring distance is unavail-
able.
E58 Lower threshold is greater than
upper threshold.
E59 Error light source reference,
video signal to dark at maximum laser
power, lamp defective or dirty.
E60 Error light source reference, mea-
sured object in the measuring range.
E61 Error light source reference,
lamp out, but raw signal to light, lamp
defective.
E65 A segment 0-0 is used for the
output or calculation.
Beschreibung
Wert für die linke Seite der Maskierung muss
kleiner sein als der rechte Wert.
Zu viele Ausgabewerte für die RS422-Schnittstel-
le ausgewählt.
Fehler bei der wiederholten Eingabe des neuen
Passwortes.
Software-Triggerung ist nicht aktiviert, es kann
kein Software-Trigger-Impuls ausgelöst werden.
Ein nicht unterstütztes Zeichen wurde empfan-
gen.
Aktuelles Signal ist nicht für die Ausgabe ausge-
wählt.
Synchronisation und Pegel- oder Flankentrigge-
rung sind nicht gleichzeitig möglich.
Das Videosignal ist zu dunkel.
Messabstand nicht verfügbar.
Untere Schwelle ist größer als obere Schwelle.
Fehler Hellabgleich, Videosignal zu dunkel
bei Laser MAximum, Lampe defekt oder ver-
schmutzt.
Fehler Hellabgleich, Messobjekt im Messbereich.
Fehler Hellabgleich, Rohsignal zu hell bei Laser
aus, Lampe defekt.
Ein nicht definiertes Segment(0-0) wird für die
Ausgabe oder Berechnung verwendet.
Seite 105