Connecting To The Power Supply; First Use; Working With The Orthopilot® 4 System; System Set-Up - B. Braun Aesculap OrthoPilot Elite FS112 Gebrauchsanweisung/Technische Beschreibung

4 panel-pc
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
5.4

Connecting to the power supply

Risk of death by electric shock!
Connect the product only to a grounded power
DANGER
supply.
Damage caused by incorrect power connection!
Use only approved power supply FS112830.
CAUTION
Use only approve power cable for power supply
FS112830.
Note
Upon assembly of the OrthoPilot
automatically supplied with power through the stand.
The power supply for the OrthoPilot
power supply, which is set up for a power input voltage of 100 V to 240 V
at a frequency of 50 Hz to 60 Hz. The mains voltage is converted into a
protective extra-low voltage of 24 V DC for the supply.
Loosen the two screws 16 and remove together with the cover 17.
Connect power supply to voltage input jack 22, see Fig. B.
Attach the cover 17 with two screws 16 to the OrthoPilot
Connect power cable to power supply.
Connect power supply to building power socket.
5.5

First use

Risk of injury and/or product malfunction due to
incorrect operation of the electromedical system!
Adhere to the instructions for use of any medi-
WARNING
cal device.
Danger to user/patient!
For license and safety reasons, any unauthorized
WARNING
modification of the operating system or the appli-
cation software is prohibited and would result in
forfeiture of the operating license.
The operating system and application software
may only be installed and maintained by
Aesculap Technical Service or by persons autho-
rized by Aesculap.
Commissioning is done by a person authorized by Aesculap and includes
the following steps:
®
Testing of the OrthoPilot
Check that all components listed on the delivery note/invoice are pres-
ent
Instruction in the use and proper handling of the OrthoPilot
tion system including provision of the country-specific instructions for
use.
Explanation and clarification of the safety instructions for use of the
®
OrthoPilot
navigation system
Attachment of the country-specific power cable
System commissioning
Installation and configuration of the required application software
Installation and configuration of the system options
Configuration of the OrthoPilot
®
®
4 stand, the OrthoPilot
®
4 panel PC FS111520 is a multi-area
to ensure perfect condition
®
4 home screen
Connection of devices
Provision of the recovery key required for data protection
Provision of the mutually signed report for employee instructions in the
use of the OrthoPilot
Data protection and data backup
Note
The hospital must ensure that, in the case of return shipment of the
OrthoPilot
and sent separately and – including when utilizing the services via the ATS
– patient data generated by the hospital during use and saved in the device
is completely and permanently deleted.
4 panel PC is
On initial commissioning and system configuration the system is
encrypted and both of the OrthoPilot
device are configured as system access dongles. One of the two
OrthoPilot
replacement key. The configured OrthoPilot
assigned to one system and cannot be used for another system. System
assignment can be tested via the identical number on the OrthoPilot
security key and the OrthoPilot
By performing this action, the system is secure from access by third par-
®
4 panel PC.
ties.
IN addition to the configured OrthoPilot
on paper is provided on first commissioning. The recovery key is required
in the case of lost or defective OrthoPilot
Pilot® USB security key can be ordered from the Aesculap Technical Service
(see Chapter 12.).
Keep the recovery key in a secure location.
6.
Working with the OrthoPilot
6.1

System set-up

6.1.1

Switching on the device

If necessary, connect the OrthoPilot
OrthoPilot
Actuate the ON-Button 1.
The button symbols light up blue.
System starts and OrthoPilot
6.1.2

Switching off the device

In order to prevent possible data loss, the OrthoPilot
properly shut down after the application software has been run through
and shut down. This is done on the OrthoPilot
Actuate the "Shutdown" button 13.
An actuation window appears.
Press the "OK" button.
®
naviga-
The OrthoPilot
ton disappears. The panel PC is now in standby mode. The OrthoPilot
wireless footswitch and the OrthoPilot
to standby mode.
Remove the OrthoPilot
protected from third-party access: remove the OrthoPilot
key from the USB port 23, see Fig. B.
®
navigation system
®
navigation system, the OrthoPilot
®
USB security keys delivered with the
®
USB security keys is required for future use. The second is a
®
4 panel PC.
®
®
USB security key to the free USB port 23, see Fig. B.
®
4 home screen 4 appears.
®
4 panel PC shuts down and the light display of the but-
®
USB security key, so that the system is securely
en
®
USB Security Key is removed
®
USB security keys are always
®
USB
USB security keys, a recovery key
®
USB security keys. A new Ortho-
®
4 system
®
USB security key: attach the
®
4 panel PC must be
®
4 home screen.
®
4 wireless camera are switched
®
USB security
®
4
7