Herunterladen Inhalt Inhalt Diese Seite drucken

Error Messages On The Orthopilot® 4 Home Screen - B. Braun Aesculap OrthoPilot Elite FS112 Gebrauchsanweisung/Technische Beschreibung

4 panel-pc
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
en
11.2 Error messages on the OrthoPilot® 4 home screen
Issue
"One of the panel PC communica-
tion ports is defective" error mes-
sage.
"The panel PC serial number is not
readable" warning message
"The panel PC is not supplied with
power/Ensure that the panel PC is
supplied with power" error mes-
sage.
"Cambar service is needed very
soon!"/"Cambar service is needed
latest on YYYY-MM-DD" warning
"The Cambar is overdue for a ser-
vice !" warning
"You have less than 4 weeks to let
your Camera be inspected!"/"Due to
service overdue, the Cambar will be
blocked on YYYY-MM-DD" warning
message
"Service is overdue for more than
one year! Please call Aesculap
Technical Service." error message.
"Cambar is warming-up. Please
wait a few minutes..." warning
"Shock sensor has been triggered.
Please call Aesculap Technical Ser-
vice." error message.
"Cambar is not functional. Please
call Aesculap Technical Service."
error message.
"Self-test has failed. Retrying..."
error message
"Cambar battery status is not avail-
able. Call Service if error persists."
error message.
"Cambar is about to go off..." error
message
14
Identification/cause
There can be no access to one of the Bluetooth
modules to connect the peripheral devices.
Software error: Panel PC serial number cannot be
read.
Failure of power supply (e.g. power cable
unplugged, power outage)
Service required (service expires on YYYY-MM-
DD)
Service urgently required. (Service interval
expired)
Service urgently required (System can no longer
be used as of YYYY-MM-DD)
Service interval exceeded. System accuracy can
no longer be guaranteed.
®
OrthoPilot
4 camera is not yet up to operating
temperature/accuracy may not be achieved.
Shock indicator triggered/Camera may be mis-
aligned.
There is a system error.
There is a system error.
USV battery status cannot be read.
Camera shuts off.
Remedy
Select the Info button and select the Reset button in the dialog
window.
If the warning message continues to appear, have the product
repaired by the Aesculap Technical Service.
Restart the system.
If the warning message continues to appear, have the product
repaired by the Aesculap Technical Service.
Connect the power cable or restore the power supply.
Have the product repaired by the Aesculap Technical Service.
Message will disappear automatically when operating temper-
ature is reached.
Have the product repaired by the Aesculap Technical Service.
Restart the system.
If the warning message continues to appear, have the product
repaired by the Aesculap Technical Service.
Restart the system.
Check the camera's power supply.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis