Note
®
If the OrthoPilot
4 panel PC cannot be properly shut down, it may be forc-
ibly shut down by pressing the ON button 1 for longer then four seconds.
This option should be used only in case of an emergency, because data may
be lost.
6.1.3

Decommissioning

Note
The safe and all-pole disconnection of the product from the main power
supply is only guaranteed when the power cord is unplugged.
Switching off the device, see Chapter 6.1.2.
Unplug the power cord from the mains power socket.
The operation of the device is safely terminated.
6.2

Connecting the accessories

Note
All interfaces behind the service cover 17 are intended solely for Service.
Service cover may only be removed by personnel authorized by
Aesculap.
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
Any equipment connected at the interfaces must demonstrably meet the
respective IEC standards (e.g. IEC 60950 for data processing equipment,
IEC/DIN EN 60601-1 for medical electrical equipment).
All configurations must comply with basic standard IEC/DIN EN 60601-1.
Any individual connecting devices with one another is responsible for such
configuration and must ensure compliance with basic standard IEC/DIN
EN 60601-1 or applicable national standards.
Note
Only USB 2.0 and USB 3.0 devices may be used.
In case of questions, contact your B. Braun/Aesculap partner or
Aesculap Technical Service, address see Chapter 12..
6.3

Functional test

Visually inspect OrthoPilot
tion.
Note
®
The OrthoPilot
4 panel PC performs a system self-test while starting which
reports any system errors, see Chapter 11..
7.

Operation

Risk of injury to the patient due to incorrectly dis-
played values!
WARNING
Displayed values may be incorrectly interpreted if
there are display faults.
If parts of the display fail or if there are pixel
errors, do not use the OrthoPilot
8
®
4 panel PC for proper mechanical condi-
®
4 panel PC.
Risk of injury and/or malfunction!
Always carry out a function test prior to each
CAUTION
use of the product.
Do not operate device if power supply cable is
defective.
Do not operate device if it has fallen and has a
defect.
Do not operate device if liquid has penetrated
into the device.
Incorrect calculations in the OrthoPilot
applications if the side to be operated on is errone-
CAUTION
ously selected!
When entering the patient data, make certain
that the correct side of operation is selected.
Failure of the device by not following the
®
OrthoPilot
CAUTION
Note the status information in the OrthoPilot
home screen.
Do not use the system if there are existing error
messages.
Note system warnings, in particular with regard
to emergency power supply/battery status.
Device failure through malware!
Check rewritable data media used to secure
reports before use using an up-to-date virus
CAUTION
scanner.
The documentation of the navigation data does not
necessarily reflect the actual medical outcome of
the operation!
CAUTION
Make certain the navigation results are regis-
tered at the correct time and with the navigated
instrument positioned correctly.
After system start, the OrthoPilot
sists of the following elements:
Patient data 7
Hospital name 8
Physician profile 9
Applications/Application processes 11
Device Status/Device Connection 4
Data security/Data cleaning 12
Settings 13
Status display (volume/date/time) 10
System shutdown 5
®
4 home screen!
®
4 home screen 6 is displayed. This con-
software
®
4

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis