11. Troubleshooting

Whenever an error message is issued by the OrthoPilot
System, shut down and restart the system, if possible.
Perform data capture again if possible.
If it is not possible to detect the error or if the error occurs again, abort
the navigated operation and complete the intervention manually.
Report error messages to the manufacturer or to one of their autho-
rized representatives, see Chapter 12.

11.1 Device Errors

Issue
Panel PC does not start after
activation of the ON switch.
The error message "COMS bat-
tery low" appears during the
panel PC boot process.
Suspected disruption/influenc-
ing of the system by a device
"BitLocker" error message at sys-
tem start-up
No touch screen input accepted
® Elite
Hardware
Identification/cause
The power cable or power supply of the
®
OrthoPilot
4 stand/external table power supply is
not connected
Defective power supply:
Defect in the power supply or power cable
Defective interface between the panel PC and
®
OrthoPilot
4 stand
CMOS battery is dead
Operation is impaired or impossible
®
OrthoPilot
USB security key not inserted into the
device
®
OrthoPilot
USB security key damaged/no longer
available
Touchscreen defective
Remedy
Connect the power cable/power supply.
Have the product repaired by the Aesculap Technical Service.
Have the product repaired by the Aesculap Technical Service.
Deactivate other devices and check whether the disruption per-
sists.
Change the spatial positioning (distance) of the devices until the
interference disappears.
Make certain that non-system cables are routed separately.
®
OrthoPilot
Shut system down (see Chapter 6.1.2) and discon-
nect it from power for approx. 10 seconds. Then restart.
Insert the OrthoPilot® USB security key into a USB port and
restart the system
Use the replacement key.
– or –
®
Reorder a OrthoPilot
USB security key via Aesculap Technical
Service. The recovery key is required for this, see Chapter 5.5.
Have the product repaired by the Aesculap Technical Service.