7.1

Patient data

In this area, preoperative patient data and data relevant to the surgery can
be entered.
Enter the patient data (name, first name, date of birth).
Select side for operation.
Select gender.
7.2

Physician profile

In this area, physician-specific profiles can be created, edited, deleted and
selected.
Note
The name of the surgeon can be entered or pre-set by selecting a profile.
7.3

Application/Application processes

Note
Applications can only be tests if all patient data is entered, see Chapter 7.1.
Select the tile for the relevant OrthoPilot
The application is started.
Note
Additional information on the applications and their workflows is available
in the instructions for use of the OrthoPilot
Configure application tiles.
Note
The edit mode must be activated to configure the application tiles, see
Chapter 7.5.
The application tiles can be configured as follows:
Enlarge/shrink tile (touch gesture)
Move tile (touch gesture)
Rename tile (entry via virtual keyboard)
Note
The edit mode must be deactivated and the user mode activated to then
start applications, see Chapter 7.5.
7.4

Device Status/Device Connection

Fig. 1
In this area, status information on the OrthoPilot
nected periphery devices OrthoPilot
wireless footswitch are displayed.
Note
®
If the OrthoPilot
4 wireless footswitch is note being used, the status dis-
play can be deactivated upon first commissioning.
Display on information on device and options for function tests and
error handling: select the button of the device.
®
application software.
®
software application.
®
4 panel PC and the con-
®
4 wireless camera and OrthoPilot
There are two types of messages:
Symbol
Designation
Warning
Malfunction
®
7.4.1
OrthoPilot
Information, such as article number, serial number and battery status of
®
the OrthoPilot
4 panel PC is displayed via the "Panel PC" button.
Battery status
If the panel PC is being powered during operation via the integrated USV,
a warning is displayed. The battery of the USV is designed so that the panel
PC can be provided with power for at least 2 minutes in the event of a
power outage. If there is a power outage during application workflow, this
will be displayed in the application (for additional information see the
instructions for use of the applications).
If the cause of the power failure cannot be remedied within 5 min, all
power supply to the system will cease when the battery status is critical.
Message
Warning: devices
is running on
emergency
power
Error: critical
battery status
Check that the mains plug has not been pulled out inadvertently.
When the OrthoPilot
matically function tested. If a bridging of voltage interruptions cannot be
guaranteed (such as due to a dead of defective USV battery), a warning is
displayed.
Only use the device when the battery has bee charge to a point that the
USV functionality can be guaranteed during operation.
®
4
Description
Device can be used under certain frame-
work conditions.
Read the warning and follow the instruc-
tions.
Device cannot be used.
Alleviate the error with the trouble-
shooting dialog.
4 panel PC
Message displayed
Device can be used under certain framework condi-
tions.
Read the warning and follow the instructions.
Device cannot be used.
Alleviate the error with the troubleshooting dialog.
® Elite
Hardware System is started up, the USV is auto-
en
9