Components
A. Top button
H. Paper backing
B. Disconnect cover
I. Adhesive
C. Insertion device
J. Tubing connector
D. Raised arrows
K. Tubing
indicating tubing
L. Site connector
direction
M. White cap for site
E. Clear window
connector
F. Cannula housing
G. Soft cannula
Instructions for use
■ Read all instructions carefully before using the
Medtronic Extended infusion set.
■ Consult the pump user guide for important
information about pump therapy prior to
connecting the infusion set. This information
includes connection and filling procedures,
possible errors, and potential risks related to
pump therapy.
■ Follow good hygienic procedures. If this infusion
set is being used for the first time, perform the
first set-up in the presence of a healthcare
4
professional.
Inserting infusion set into body
Unfold the front cover of this booklet to view
images that are keyed to the steps below.
1.
Wash hands with soap and water.
2.
Select a recommended insertion site (shown
in gray) and as indicated by a healthcare
professional.
3.
Clean the insertion site with a disinfectant
as directed by a healthcare professional.
Allow to air dry before inserting the infusion
set. Remove hair around the insertion site
to ensure the adhesive sticks to the skin if
needed.
4.
Open the package and remove the paper
from the tubing.
5.
Place the tubing connector on the top of a
newly filled reservoir. If a MiniMed tubing
connector is in use, twist the tubing connector
clockwise until it locks in place.
For the MiniMed tubing connector: If insulin
or any other liquid gets inside the tubing
connector, it can temporarily block the vents
that allow the pump to properly fill the infusion
set. This may result in the delivery of too
little or too much insulin, which can cause
hyperglycemia or hypoglycemia. If this
occurs, start over with a new reservoir and
infusion set.
6.
Remove the white cap from the site connector
by (a) squeezing the sides of the site
connector then (b) pulling the white cap off.
Keep the white cap for later use.
7.
Put the reservoir into the pump, then
completely fill the tubing with insulin as
instructed by the pump user guide. Do not
leave any air in the tubing. The tubing is
completely filled when drops of insulin are
seen coming out of the site connector.
8.
Remove the paper backing from the
adhesive. Be careful not to touch the
adhesive.
9.
Remove the disconnect cover from the
insertion device by gently squeezing the
sides of the disconnect cover then pulling
it away from the insertion device. Keep the
disconnect cover for later use. The raised
arrows indicate the tubing direction.
10a. Stretch the skin until smooth. Then press the
insertion device against the skin.
10b. Press the top button down completely to
insert the infusion set.
11. Gently and carefully remove the insertion
device.
12. Gently press the adhesive onto the skin with
one finger. Replace the infusion set if the
adhesive does not stick to the skin.
13. Gently hold the cannula housing steady with
one finger. Then push the site connector
straight into the cannula housing until it clicks.
14. Fill the soft cannula with insulin:
6 mm: 0.6 units (0.006 ml)
9 mm: 0.6 units (0.006 ml)
Dispose of the insertion device in an appropriate
sharps container according to local laws.
Keep the disconnect cover and the white cap for
use when the infusion set is disconnected.