Once all detergent residue is removed, continue to rinse for 30 seconds.
Drain excess water from the device and dry it using a clean cloth or
pressurized air.
Visually inspect each device for cleanliness, paying close attention to hard-to-
reach areas (see Figure 56).
1 Reverse Osmosis/Deionized water (RO/DI) at 30°C was used for validating cleaning efficacy.
2 Neodisher MediClean Forte will work at a ratio of 5-30 ml/L.
Figure 56: Hard to reach areas on the PINPOINT camera
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
93