If the PINPOINT system fails to produce continuous imaging, switch the main
power off (on the VPI rear panel) for five seconds and then switch the power on
again.
The PINPOINT camera, light guide cable, and camera cable should remain
connected to the VPI during this time.
If fluorescence imaging fails to resume:
56
Disconnect the laparoscope from the PINPOINT camera.
Use direct visualization through the laparoscope eyepiece in conjunction with
white light illumination to safely conclude the procedure.