WARNING:
Proper sterile technique must be used when preparing the camera and
laparoscope for surgical endoscopy. The following components are not
sterile and should not be handled by a sterile operator:
Video Processor / Illuminator (VPI)
Camera Connector
Do not remove the cap of connector while it is in a sterile field.
To prepare the PINPOINT camera and laparoscope:
Inspect the laparoscope prior to connecting it to the camera:
▪ Ensure that no parts are missing or loose.
▪ Ensure that the proximal end is dry to prevent the laparoscope from
▪ Ensure that there are no residual cleaning agents or disinfectants on the
▪ Inspect the entire laparoscope, particularly the shaft, for contaminants and
▪ Inspect distal end, proximal end and irradiation surface of the illumination
Connect the desired laparoscope to the camera following the instructions in
Chapter 2.
Note:
High-Definition Imaging Laparoscopes are optimized for imaging with visible light
and will not perform ICG fluorescence imaging. To perform ICG fluorescence
imaging, use an ICG Fluorescence Imaging Laparoscope.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
fogging during examination / procedure.
laparoscope.
damage of any type, such as dents, scratches, cracks, bending or sharp
edges.
fibers for contamination and scratches. Make contaminants and scratches
visible using light reflections. Hold the connection of the optical fibers
against the light and inspect whether the optical fibers illuminate evenly at
the distal end. Contaminants on the irradiation surface of the illumination
fibers can damage fibers, which impacts image quality.
39