Reprocessing the PINPOINT camera requires a four-step process:
Step 1
Step 2
Step 3
Step 4
WARNING:
Following all four steps is critical to ensuring the component is ready for
use. Failure to properly prepare and clean the device could result in a
contaminated component after reprocessing.
When preparing and cleaning, ensure all foreign matter is removed from the
surface of the device. This allows the active ingredients of the sterilization method
to reach all the surfaces of the device.
CAUTION:
Do not clean the PINPOINT camera using an ultrasonic bath.
CAUTION:
Do not use metal or abrasive brushes, scrub pads, or rigid tools when cleaning the
PINPOINT camera. The optical elements can be scratched, permanently damaging
the equipment.
Note:
The PINPOINT camera may be contaminated by contact with human tissue and
fluids during clinical use. To avoid drying of blood, protein, and other substances
on the camera, prepare it for cleaning within one hour after use.
Detailed pre-treatment instructions:
92
Preparation
Manual Cleaning
Inspection
Sterilization
Fill a wash basin with lukewarm water.
Measure and dispense enzymatic detergent into the water based on the
detergent manufacturer instructions.
Gently mix the detergent into the water by hand.
Submerge the device into the prepared wash basin.
With the device immersed in the solution, thoroughly brush the exterior with a
soft-bristled brush, focusing on any mated or rough surfaces.
Actuate and brush any movable parts in all extreme positions.
Rinse each device with water
4-0002428 Rev. J
1
2
1
until all detergent residue is removed.