Reprocessing laparoscopes and light guide cables requires a four-step process:
Step 1
Step 2
Step 3
Step 4
WARNING:
Following all four steps is critical to ensuring the component is ready for
use. Failure to properly prepare and clean the devices could result in
contaminated components after reprocessing.
When preparing and cleaning, ensure all foreign matter is removed from the
surface of the device. This allows the active ingredients of the sterilization method
to reach all the surfaces of the device.
CAUTION:
Laparoscope optics may be damaged if the instruments are dropped. Always
handle with care.
Handle the light guide cable with care as damage to the protective outer cover
causes damage to fiber optic light cable. Protect the light guide cable from bending
tightly which may damage the inner fiber bundle. It is very important to prevent
impacts, especially to the fused light entry, as this can lead to the destruction of the
fusing.
CAUTION:
Do not clean laparoscopes or light guide cables using an ultrasonic bath.
Use only approved cleaning agents. Do not use alcohol or other corrosive cleaning
agents.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
Prepare
Clean
Inspect
Sterilize