Visually inspect each device for cleanliness, paying close attention to hard-to-
reach areas.
1 Reverse Osmosis/Deionized water (RO/DI) at 30°C was used for validating cleaning efficacy.
2 Neodisher MediClean Forte will work at a ratio of 5-30 ml/L.
3 Neodisher MediClean Forte recommends a soaking time of 5 minutes.
Detailed automated cleaning instructions:
Place the device in the automated washer on an incline to facilitate drainage.
Program the washer using the parameters in Table 9:
Table 9: Automated cleaning parameters
Pre-Wash
Wash 1
Rinse 1
Dry Phase
Thermal Rinse
Drain excess water from the device and dry it using a clean cloth or
pressurized air.
Visually inspect each device for cleanliness, paying close attention to hard-to-
reach areas.
1 Neodisher MediClean Forte is the validated detergent and will work at a ratio of 5-30 ml/L.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
2 minutes
5 minutes
2 minutes
2 minutes
1 minute
Cold water
Set Point 60°C (140°F)
Hot Water
115°C (239°F)
90°C (194°F)
N/A
1
Enzymatic
N/A
N/A
N/A