WARNING:
Grounding reliability can only be achieved when equipment is connected to
an equivalent receptacle marked 'hospital only' or 'hospital grade'.
Connect the PINPOINT system components only to approved medical
systems or to systems that are powered from approved isolation
transformers.
The patient leakage currents from multiple, simultaneously used, energized
endoscopic accessories may be additive. Use PINPOINT only with Type CF
Applied Part energized endoscopic accessories that minimize leakage
currents to the patient.
Avoid the use of PINPOINT with high-frequency (HF) energized endotherapy
devices in the event of explosive gas concentrations being present in the
area of use.
Do not use the imaging system if the power cord or plug is damaged or
modified in any way.
Do not remove or override the ground connection on the power cords.
CAUTION:
Unplug power cords by grasping the plug. Do not unplug power cords by pulling on
the cable.
WARNING:
Avoid looking at light emitted directly from the laparoscope or the light guide
cable tip. Table 19 in Appendix C contains specifications for NIR radiation
emitted in fluorescence mode.
Use of controls or performance of procedures other than those specified
herein may result in hazardous radiation exposure.
Always use standby mode by turning illumination off (see Turning Off
System Illumination, in Chapter 5) when connecting or disconnecting the
light guide cable from the laparoscope, or prior to removing the laparoscope
from the patient.
There is high-energy light at the distal end of the laparoscope. This can
cause the temperature of the body tissue to rise to 41°C. Avoid direct contact
of the distal end with body tissue or flammable materials as it can cause
burns.
A laser aperture label is located on the surface of the PINPOINT camera as
prescribed by applicable standards (see Figure 2).