WARNING:
The PINPOINT System has been certified for compliance with international
standards for electromagnetic compatibility (EMC).
The PINPOINT system generates radio frequency energy and should be
installed and used in accordance with the instructions to minimize the
possibility of interference with other electro-medical equipment. However,
there is no guarantee that interference will not occur in any particular
installation.
Please see Table 12 in Appendix A if the PINPOINT System is suspected of
causing interference with other electro-medical equipment, despite being
installed according to the instructions.
Read and follow the information in "Electrical Safety – Power" in Chapter 1
WARNING:
and only connect the PINPOINT VPI to a "Hospital Only" or "Hospital Grade"
power outlet.
Select a location in which to use the PINPOINT System that is within 3 m
access of an appropriate power outlet.
Place the VPI on an endoscopy cart, on a shelf supported by a ceiling-mounted
boom, or on a suitable table.
Select a location in which to use the PINPOINT System that is within 3 m
access of a medical-grade color video monitor.
CAUTION:
Ensure that the selected location provides a minimum 5 cm (2 inches) gap around
the VPI cooling vents to allow for required ventilation.
Do not place heavy objects directly on top of the VPI.
For more information, please refer to Table 14 in Appendix C.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
19