After proper cleaning and inspection, laparoscopes and light guide cables may be
sterilized using NOVADAQ Sterilization Trays (PC9017).
WARNING:
Use only legally-marketed autoclaves. All equipment used must be legally
marketed in accordance with the governing regulatory body.
WARNING:
Laparoscopes must be sterilized in the sterilization tray supplied by
NOVADAQ.
Maintain sterilization trays regularly and per manufacturer's instructions.
WARNING:
DO NOT Use "Immediate Use" autoclave sterilization (formerly known as
WARNING:
flash sterilization).
Follow sterilizer manufacturer's instructions.
WARNING:
WARNING:
Relevant, national/regional legal regulations must be observed.
CAUTION:
Ensure that optical surfaces do not contact hot metal surfaces, as this may damage
the entire system.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
87