6.3.1

Activating the motor/pump control panel

Note
Motor settings of applied parts that are not connected to a cable cannot be
changed while the applied part is running.
Changing the setting parameters of the applied part: Activate the
applied part control panel b in the display 2, see Fig. 5.
Changing the setting parameters of the coolant and irrigation pump:
Activate the pump control panel a in the display 2.
The activated control panel switches to the Settings mode. Now the
setting parameters listed below can be changed.
b
a
Fig. 5
Activating the control panel
Legend
a
Pump control panel
b
Applied part control panel
6.3.2

Changing the maximum speed setting/rotational direction of

the applied part

Activate the applied part control panel, see Chapter 6.3.1.
Change direction of rotation (except for drill tool GA844): Actuate the
inactive, grey clockwise/counterclockwise rotation button a/b, see
Fig. 6.
The rotational direction is changed from clockwise rotation to counter-
clockwise rotation and vice versa.
Change speed range limit: Press the buttons to reduce/increase the
speed range limit c/d or the speed range limit switch 31 on the wireless
foot switch.
The maximum speed setting is reduced/increased by increments.
Note
The increments for changing the maximum speed setting depend on the
applied part connected.
a
b
c
d
Fig. 6
Changing the maximum speed setting/rotational direction
12
Legend
a
Rotational direction button for counterclockwise rotation
b
Rotational direction button for clockwise rotation
c
Button for reducing the maximum speed setting
d
Button for increasing the maximum speed setting
6.3.3
Activating the coolant and irrigation pump/"Flush" function
(continuous flushing)
Activate the pump control panel, see Chapter 6.3.1.
Activate the coolant and irrigation pump: Activate the button for
switching on the irrigation pump a, see Fig. 7.
The coolant and irrigation pump is active and delivers coolant or irri-
gation fluid at the currently set delivery rate.
Activate the "Flush" function: Activate and hold down the "Flush"
button b.
The "Flush" function is now active. The coolant and irrigation pump
delivers coolant or irrigation fluid at the maximum delivery rate until
the "Flush" button b is released.
Note
The activation of the "Flush" function does not depend on whether an
applied part is connected or activated.
a
b
c
d
Fig. 7
Activating the coolant and irrigation pump/"Flush" function
Legend
a
Button for switching the coolant and irrigation pump on/off
b
"Flush" button
c
Button for reducing the delivery rate
d
Button for increasing the delivery rate
6.3.4
Deactivating the coolant and irrigation pump and changing
the delivery rate
Note
The delivery rate of the coolant and irrigation pump can only be changed
when the irrigation pump is activated.
Activate the pump control panel, see Chapter 6.3.1.
Deactivate the coolant and irrigation pump: Activate the button for
switching off the irrigation pump a, see Fig. 7.
The coolant and irrigation pump is inactive and no longer delivers cool-
ant.
Change the delivery rate: Activate the buttons for reducing/increasing
the delivery rate c/d.
The delivery rate of the coolant and irrigation pump is
reduced/increased by increments.
The delivery rate can be adjusted in the following increments:
1 % to 5 %: 1 % increments
5 % to 100 %: 5 % increments