Before attempting the insertion, ensure that you are familiar
with the potential complications and their emergency treatment
should any of them occur.
WARNINGS:
In the rare event that a hub or connector separates from any
component during insertion or use, take all necessary steps and
precautions to prevent blood loss or air embolism and remove
catheter.
Do not advance the guidewire or catheter if unusual resistance
is encountered.
Do not insert or withdraw the guidewire forcibly from any
component. The wire may break or unravel. If the guidewire
becomes damaged, the introducer needle or Vascu-Sheath
introducer and guidewire must be removed together.
Federal Law (USA) restricts the device to sale by or on the order
of a physician.
This catheter is for Single Use Only.
Do not resterilize the catheter or accessories by any method.
Re-use may lead to infection or illness/injury.
The manufacturer shall not be liable for any damages caused by
re-use or resterilization of this catheter or accessories.
Contents sterile and non-pyrogenic in unopened, undamaged
package.
STERILIZED BY ETHYLENE OXIDE
Do not use catheter or accessories if package is opened or
damaged.
Do not use catheter or accessories if any sign of product
damage is visible.
CATHETER PRECAUTIONS:
Do not use sharp instruments near the extension tubing or
catheter lumen.
Do not use scissors to remove dressing.
Catheter will be damaged if clamps other than what is provided
with this kit are used.
Clamping of the tubing repeatedly in the same location may
weaken tubing. Avoid clamping near the luers and hub of the
catheter.
Examine catheter lumen and extensions before and after each
treatment for damage.
To prevent accidents, assure the security of all caps and
bloodline connections prior to and between treatments.
Use only Luer Lock (threaded) Connectors with this catheter.
Repeated over tightening of bloodlines, syringes, and caps will
reduce connector life and could lead to potential connector
failure.
STERILE EO
-2-