Prepare for Testing
The following steps assume that users of a BD Veritor Plus Analyzer have chosen and set all configuration options, and
that the Analyzer is ready to use. To choose or change these settings, see the BD Veritor Plus Analyzer Instructions for
Use, section 4.7. A printer is not necessary to display results. However, if your facility has chosen to connect the Analyzer
to a printer, check that the printer is plugged into a power source, paper supply is adequate and any necessary network
connections are enabled before testing.
Step 1: For each patient specimen:
• Remove one RV Reagent D tube/tip and one BD Veritor System Flu A+B device from
its foil pouch immediately before testing.
• Label with patient's name or ID number.
• Place the labeled RV Reagent D tube(s) in the designated area of the tube rack.
Prepare the Sample
Step 2:
• Remove and discard the cap from the RV Reagent D tube corresponding to the
sample to be tested.
Step 3:
• Insert the patient sample swab all the way into the RV Reagent D tube and swirl it
against the inside wall three (3) times.
Step 4:
• Remove the swab while squeezing the sides of the tube to extract the liquid from
the swab. Properly discard of the swab.
Step 5:
• Press the attached tip firmly onto the RV Reagent D tube containing the processed
sample (threading/twisting not required).
• Vortex or mix thoroughly by swirling or flicking the bottom of the tube before adding
to assay device.
• Note: Do not use tips from any other product, including other products from BD or
other manufacturers.
After step 5, choose from the model and workflow option below before continuing to step 6:
Instructions in
section:
BD Veritor Reader or
Analyzer in
Analyze Now mode
A
BD Veritor Plus
Analyzer in
Walk Away mode
B
5
BD Veritor Plus Analyzer with the
BD Veritor InfoScan module in
Analyze Now mode–or–Walk Away mode
C
D