56 | ROLL R6 M | Commissioning
4.5

Electrical connection

DANGER – Risk of electric shock if
connections are not or not correctly
grounded or from incorrect mains voltage.
Incorrect phase, neutral and ground
connections can cause electric shocks, heart
failure and fatalities!
Work on electrical installations or
equipment is only to be performed by
qualified electricians or trained personnel
under the guidance and supervision of an
electrician.
Even minor work on the electrical system
is only to be performed by appropriately
trained and qualified personnel.
The ROLL R6 M is only to be connected
to the power supply if the mains voltage
available corresponds to the rated voltage
given on the rating plate.
Check the grounding before connecting
the ROLL R6 M.
Use the type of power cord specified for
the country concerned.
Only connect to an appropriate approved
grounding-type socket.
In the event of operating problems,
switch off immediately by way of the On/
Off switch, disconnect the power cord
and consult the "Faults" section of the
operating instructions.
i
The ROLL R6 M must be connected to a standard
electrical system provided with a residual-current-
operated circuit breaker with a connection cross-
section of at least 3 mm conforming to European
standards. The customer is responsible for providing
fuse protection for the mains connection.
i
The ROLL R6 M is connected to the mains by way of
a socket. For this purpose use is to be made of the
detachable cable with standard connector supplied
with the device. The connector must always be
readily accessible.
i
A two-pole residual-current-operated circuit breaker
of type "C" must always be used to provide fuse
protection for the mains connection. Single-pole
circuit breakers are not permissible.
|
1 695 600 694
2016-09-28
i
The temperature and ambient conditions specified
in the Technical Data must be observed for
operation.
Fig. 23:
Power cord connection - Back of ROLL R6 M
1 On/off switch
2 Mains connection
3 Power cord