Cursor measure position menu as below:
Function
Display
M1-M2
Position
Cursor 1
Position
Cursor 2
Position
Follow up below steps to measure data width of pulse with cursor measurement
1.
Press "9(CURSOR)" and Cursor menu appears.
2.
Press "F1" till cursor display as "Position". Two purples vertical lines display in panel and refer to Cursor
1 and Cursor 2 separately.
3.
Turn "CH1 POSITION" knob and set Cursor 1 position in falling edge of CH00
Turn "CH2 POSITION" knob and set Cursor 2 position in next falling edge of CH00.
4.
5.
Now the position difference between two cursors display in "F2" is "-10" which means there is 10 acquired
data between two cursors. The position of Cursor 1 corresponds to trigger in memory position display in
"F3" as "2097158". The position of Cursor 2 corresponds to trigger in memory position display in "F4" as
"2057168".
Fig. 102 Cursor time measurement
Instruction
The position difference between two cursors in memory area.
The position of cursor 1 corresponds to trigger in memory area.
The position of cursor 2 corresponds to trigger in memory area.