Herunterladen Inhalt Inhalt Diese Seite drucken

PeakTech 1190 Bedienungsanleitung Seite 177

Digital oszilloskop mit 16-kanal-logik-analysator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

200 kHz
100 kHz
Follow up steps to set the measurement for data bus of 33M clock:
Sample rate: 100M Storage depth: 4M
Filter: none
Setting steps as below:
1. Press "E(ACQUIRE)" and sample menu appears.
2. Press "F1"repeatly or turn "CH1 Volts/div" knob till sample rate set as "100M"
3. Press "F2" till storage depth display as "Deep Memory".
4. Press "F3" till digital filter display as "None".
Then sample system setting finished (refer to Fig. 75).
44.2. How to set trigger system
Trigger system including "Trigger level" adjust knob, "Trigger menu", "SET 50%", "Set to Zero". Trig level
adjust knob is to adjust trigger position. Since the digital storage depth is fixed so the proportion of storage
data between pre-trigger points and re-trigger ones depend on trigger position. Trigger mode including: Edge
trigger, bus trigger, pattern trigger, sequential queue trigger, distributed queue trigger.
Low memory
80 ms
General
1.28 s
Deep memory
20 s
Low memory
160 ms
General
2.56 s
Deep memory
40 s
Fig. 75 Sampling setting
Low memory
20 Hz
General
Deep memory
-176-
800 s
12800 s
200000 s