33. XY Format
This format is only applicable to Channel 1 and Channel 2. After the XY display format is selected, Channel 1
is displayed in the horizontal axis and Channel 2 in the vertical axis; the oscilloscope is set in the un-triggered
sample mode: the data are displayed as bright spots and the sampling rate is 1MS/s and cannot be changed.
The operations of all control knobs are as follows:
*
The Vertical VOLTS/DIV and the Vertical POSITION knobs of Channel 1 are used to set the horizontal
scale and position.
*
The Vertical VOLTS/DIV and the Vertical POSITION knobs of Channel 2 are used to set the vertical
scale and position continuously.
The following functions cannot work in the XY Format:
*
Reference or digital wave form
*
Cursor
*
Auto Set
*
Time base control
*
Trigger control
Operation steps:
Press the DISPLAY button and call out the Display Set Menu.
1.
2.
Press the F3 menu selection button and choose the form as XY. The display format is changed to be XY
mode (see Fig.56).
Fig. 56 XY Display Mode
-160-