44.5. How to set BUS
LA includes four groups BUS (BUS0~BUS3). Every group can cover any channel even all channels. BUS
menu display as Fig. 92.
BUS setting menu as below:
Function
Setting
BUS
BUS0~BUS3
CH0F~CH00
Channel
1X111111
11111111
Include
CH0F~CH00
Exclude
Code
HEX
Type
DEC
For example: to set BUS0 as BUS to include CH00, CH01, CH02, CH03 and code as hex system. Follow up
as below:
1.
Press "0(BUS)" to and BUS menu appears
2.
Press "F1" till BUS display as "BUS0"
3.
Press "F2" or turn "CH1 Volts/div" knob till channel display as "CH00"
4.
Press "F3" and set CH00 as "Include". Then Channel CH00 refer to display of "1"
5.
Repeat the operation of steps 2. 3 and set CH01, CH02, CH03 as "Include" and other channel as
"exclude". Then "CH SEL" display as "XXXXXXXXXXXX1111". BUS setting finished.
6.
Press "F4" and set code type as "HEX"
Fig. 92 Bus menu
Instruction
Choose BUS for operating
Choose any channel among CH00~CH0F
BUS channel complex indication: 1 for include; X for exclude
The bus selected includes this channel
The bus selected excludes this channel.
All of the code type and measured value are HEX
All of the code type and measured value are DEC