7.1 Electromagnetic Compatibility IEC/EN 60601-1-2
WARNING: Contraindications.
Interference with other equipment.
Although PIEZOSURGERY plus complies
with standard IEC/EN 60601-1-2, it may
interfere with other equipment located
nearby. PIEZOSURGERY plus, with relative
accessories, must not be used near, or
stacked on other equipment. Install
PIEZOSURGERY plus at a safe distance
from life-support equipment. If necessary,
however, prior their use, please check
and monitor that the device, the relative
accessories and all other equipment works
correctly in that layout.
WARNING: Portable and mobile
radiocommunications equipment can
influence the correct operation of the
device and its accessories.
7.2 Guidance and Manufacturer's Declaration -
Electromagnetic Emissions
PIEZOSURGERY plus, with relative accessories, is designed to operate in the electromagnetic
environment specified below. The customer or the user of PIEZOSURGERY plus should ensure
that it is used in such environment.
Emissions Test
RF Emissions
CISPR 11
36
Interference by other equipment.
An electrosurgical knife of other
electrosurgical devices near the
PIEZOSURGERY plus device and to its
relative accessories, may interfere with its
correct operation.
with relative accessories, requires specific
EMC precautions and must be installed
and commissioned according to the EMC
information provided in this chapter.
components not supplied by Mectron might
negatively affect the EMC performances.
Compliance
Electromagnetic Environment Guidance
PIEZOSURGERY plus uses RF energy
only for its internal operation. The
PIEZOSURGERY plus could emit frequencies
causing interference with other devices in
Group 1
the vicinity.
Install the PIEZOSURGERY plus away from
life-support equipment. Check correct
operation of all the equipment before
starting the operation.
WARNING: Contraindications.
WARNING: PIEZOSURGERY plus,
WARNING: The use of cables and