Fuses
Operating frequency
Power settings
Peristaltic pump capacity
APC circuit protections
Malfunction detection system
Operating conditions
Transport and storage conditions
Altitude
Weight and size
a) W = width ; L = length ; H = height
TECHNICAL DATA
Type 5 x 20 mm, T 5AL, 250V
Automatic scan
from 24 KHz to 36 KHz
Adjustable on the touch screen:
7 power levels, from 1 to 7.
NOTE: For each insert, the minimum
and maximum power levels are preset
by Mectron.
Configurable via the touch screen:
5 flow levels, from a minimum of 4 ml/
min to a maximum of approximately 81
ml/min.
NOTE: For each insert, the minimum
and maximum irrigation levels are
preset by Mectron.
No handpiece detected
Cord interruption
Insert not tightened correctly or broken
The system displays an anomaly detected
and the corrective action to be taken (see
Chapter 8 on page 44).
from 10 °C to 30 °C
Relative humidity from 30% to 75%
Air pressure P: 800hPa/1060hPa
from -10 °C to 60 °C
Relative humidity from 10% to 90%
Air pressure P: 500hPa/1060hPa
Less than or equal to 2000 meters
11,7 Kg
380 x 410 x 310 mm (W x L x H)
)
a)
35