mectron Piezosurgery plus Gebrauchs- Und Wartungsanleitung Seite 51

Vorschau ausblenden Andere Handbücher für Piezosurgery plus:
Inhaltsverzeichnis
Basic EMC
Phenomenon
standard or test
Electrical fast
transients /
IEC 61000-4-4
bursts
l) o)
Surges
b) j) o)
IEC 61000-4-5
Line-to-line
Surges
b) j) k) o)
Line-to-
IEC 61000-4-5
ground
Conducted
disturbances
IEC 61000-4-6
induced by RF
fields
c) d) o)
Voltage dips
IEC 61000-4-11
f) p) r)
Voltage
interruptions
IEC 61000-4-11
f) i) o)
a) Void.
b) All PIEZOSURGERY plus cables are attached during the
test.
c) Calibration for current injection clamps shall be
performed in a 150 Ω system.
d) If the frequency stepping skips over an ISM or amateur
band, as applicable, an additional test frequency shall be
used in the ISM or amateur radio band. This applies to
each ISM and amateur radio band within the specified
frequency range.
e) Testing may be performed at other modulation
frequencies identified by the RISK MANAGEMENT
PROCESS.
Immunity test levels
method
±2 kV on contact
100 KHz repetition
frequency
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2kV
3 V
m)
0.15 MHz - 80 MHz
6 V
in the ISM bands
m)
between 0.15 MHz and
80 MHz
80% AM at 1 KHz
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
q)
0% UT; 1 cycle e
70% UT; 25/30 cycle
Single phase: at 0°
0% UT; 250/300 cycle
TECHNICAL DATA
Environment Guidance
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Portable and mobile RF
communication devices
should not be used near any
part of the product, including
cables, except when they
respect the recommended
n)
distances, calculated from the
e)
equation applicable at the
frequency of the transmitter.
g)
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
h)
The quality of the network
voltage should be that of a
h)
typical commercial or hospital
environment.
f) ME EQUIPMENT and ME SYSTEMS with a d.c. power
input intended for use with a.c.-to-d.c. converters shall
be tested using a converter that meets the specifications
of the MANUFACTURER of the ME EQUIPMENT or ME
SYSTEMS. The IMMUNITY TEST LEVELS are applied to
the a.c. power input of the converter.
g) Applicable only to PIEZOSURGERY plus connected to
single-phase a.c. mains.
h) E.g. 10/12 means 10 periods at 50 Hz or 12 periods at
60 Hz.
Electromagnetic
39