NOTE: the foot pedal is provided with
a bracket that, if positioned vertically,
enables it to be moved to the most
appropriate place for use, with no need
to touch it with hands.
The bracket can also be positioned
horizontally, if not in use;
CAUTION: The user must pay
particular attention to the foot pedal
positioning, so that the pedal can
only be activated intentionally by the
operator.
Insert the solution bag stand into the
proper holes;
Install the two peristaltic pumps, gently
pushing the head of the peristaltic pump
towards the corresponding base (on the
device). Carefully rotate the pump head
forwards and backwards to ensure the
two parts match;
Rotate the pump in a clockwise direction
until the head locks into position (with a
clicking sound) and is arranged vertically.
Pull the pump gently to ensure that it is
correctly positioned.
14
4
5
6
7
CLICK!