area. During the intervention, regularly
check the insert integrity, in particular on
the apical part. During the intervention,
avoid prolonged contact with retractors or
other metal tools being used.
WARNING: The PIEZOSURGERY plus
device is intended for bone surgery.
However, prolonged contact and/or
excessive force of the instrument tip on soft
tissues should be avoided as this may cause
thermal and/or blunt injury. Take particular
care when sharp inserts are used. Prolonged
mechanical action of a sharp insert may also
result in the soft tissue being cut. In close
proximity to soft tissues/nerves (e.g. the
perinevrium of the peripheral nerve system,
or dura mater of central nervesystem),
it is recommended to complete the cut
with a blunt tip diamond coated insert,
to minimize the potential for soft tissue
damage.
WARNING: Before using the
PIEZOSURGERY medical handpiece on the
patient, check that all parts of the device are
properly working.
CAUTION: Intermittent function.
Prolonged use can cause the handpiece
INSTALLATION AND CONFIGURATION
to overheat. Refer to
34
for average use times (intermittent
function).
WARNING: Sterile single-use. Sterile
single-use objects must only be used in a
single surgical operation and for a patient,
only. Single-use objects must not be reused.
Sterile items cannot be used past their
expiration date. Separate and dispose of
each single-use object according to current
regulations governing hospital waste.
WARNING: During surgery, do not
perform any maintenance activity on the
system.
WARNING: PIEZOSURGERY plus
with accessories and compatible inserts
is classified for intermittent mode of
operation (as reported in
on page 34
PIEZOSURGERY plus with accessories and
compatible inserts for a prolonged time and
in any case exceeding the declared time
limits can cause overheating, in particular, of
the handpiece. In case of overheating, avoid
contact of the handpiece with the operator
and the patient.
Chapter 7 on page
Chapter 7
). The continuous use of
31