Error Message
MED - 050 a MED -058
Internal problem.
"SHUTDOWN" the device
then start it again.
MED - 063
The irrigation tubing line
shall be filled by pressing
the PUMP button before
using the handpiece.
MED - 065
Handpiece not correct,
change channel or
handpiece.
NOTE: For any error message not listed in
the above Table, contact a Mectron Service
Centre.
Description
Internal problem.
Filling cycle performed for
a time shorter than the
minimum required. The
filling cycle that follows
can be interrupted by the
operator.
Handpiece connected, but
not suitable for the channel
or not supported by the
system.
TROUBLESHOOTING
Actions Required
"SHUTDOWN" the device
and then start it again. If the
problem persists, contact
MECTRON assistance.
Press the "PUMP" button to
fill the irrigation circuit.
Adopt the handpiece
suitable for the selected
channel.
47