Emissions Test
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Emissions of fluctuations
voltage/flicker
IEC 61000-3-3
7.3 Accessible Parts of the Casing
PIEZOSURGERY plus, with relative accessories, is designed to operate in the electromagnetic
environment specified below.
The customer or the user of PIEZOSURGERY plus should ensure that it is used in such
environment.
Phenomenon
standard or test
Electrostatic
IEC 61000-4-2
discharge (ESD)
Compliance
Class A
Class A
Compliant
Basic EMC
Immunity test levels
method
±8 kV on contact
±2 kV, ±4 kV, ±8 kV,
±15 kV in air
TECHNICAL DATA
Electromagnetic Environment Guidance
The PIEZOSURGERY plus is suitable for
use in all establishments other than
domestic, and may be used in domestic
establishments and those directly connected
to the public low voltage power supply
network that supplies buildings used for
domestic purposes, provided the following
warning is heeded.
CAUTION: This device is intended for
use by healthcare professionals only.
The EMISSIONS characteristics of the
PIEZOSURGERY plus make it suitable for
use in industrial areas and hospitals (CISPR
11 class A). If it is used in a residential
environment (for which CISPR 11 class B is
normally required) the PIEZOSURGERY plus
might not offer adequate protection
to radio-frequency communication
services. The user might need to take
mitigation measures , such as relocating or
reorienting the PIEZOSURGERY plus.
Electromagnetic
Environment Guidance
The floor must be made of
wood, concrete or ceramic
tiles. If the floor is covered
with synthetic material, the
relative humidity should be at
least equal to 30%.
37